Enhanced Parallel Path PTFE Nebulizer, can operate on 0.2 ml/min to 2.5 ml/min. Excellent mist, and can handle high level of particulates in sample. Fits regular chambers as a direct replacement for glass concentric nebulizers. Operates on standard pressures of 35 - 45 psi.
Please specify your instrument type when ordering. Place in the comments section of the Checkout screen. This will allow us to add the proper fitting kit with your order.
The Teflon Mira Mist can be used for any liquid. Salting is very rare, but can occur over long periods of time, especially with high Sodium levels in the solution. Suitable for high solids applications such as environmental, geological, etc. Such salting can be cleared easily with a dip in 5 - 10% HF.
Caution: Be careful when handling HF. It soaks into the skin before you feel it burning, and in low concentrations it looks like water.
Cross Reference
| Agilent | G3161-80000, CP913246 |
| PerkinElmer | N0777031 |
| Thermo Scientific | 4600356 |
All Burgener nebulizers are sold with a 3 month total satisfaction guarantee (starting from the date they were delivered). If for any reason there is a problem with a nebulizer you have received within the first 3 months, we will send a replacement free of charge immediately.
The original nebulizer will need to be returned after you have received the replacement so they can examine the nebulizer and identify the problem.
Please Note: The nebulizer exterior is very durable, but the tip of the nebulizer is extremely delicate and can be easily damaged. This warranty is meant to cover manufacturing defects and does not apply if the nebulizer has been obviously mishandled and physically damaged ie. has been dropped on the floor without the protective cap in place.